Industrial Precision Instruments’ probes are formulated to specifically meet our customer’s needs, ensuring longevity, performance, accuracy and reliability with minimum maintenance. The design specifications are included in our part numbers. To find out what your part number means, refer to the list below.

Part Number Meaning
PN Part Number
C 30% Solid KCL Polymerised into Porous Kynar Cell (Liquid Free Reference)
TJ Triple Junction Reference to give extended life (Reduce poisoning of Ref)
HR Hysteresis Resistant (Reduces the problem of Tempt cycling of Ag / AgCl ratio)
6000 The 6000 series tells us the probe body is High Tempy ULTEM plastic
68 ORP Probe
64 The ULTEM body has an Acid Resistant Dome Glass (Almost unbreakable)
66 The ULTEM body with Sulphide Resistant Ref (Stop poisoning – longer probe life)
61 The ULTEM body with High Temperature rating 135C (Sealants & components eg. Preamp)
5081 Model of the TRANSMITTER – including correct wiring code
25 Length of the Signal Cable in feet
WPB Extra Water Proofing on the back end of the Probe to stop attack on the cable
TS Teflon filled Silicon to stop Acid @ 100C attacking the Probe.